(3) Automatic sensitivity registration by receiving command
By connecting the WGA-670B to the PC using RS-232C interface, the automatic sensitivity registration
command is received via RS-232C interface from the PC
For details of PC connection, see "7. RS-232C OPTION."
The command of the automatic sensitivity registration is described in the following.
Command:
Parameter:
Return value:
(4) When error is displayed
Owing to the following cause, an error message 'E-03' may appear when conducting the automatic
sensitivity registration. At this time, no sensitivity is registered.
Pressing the CANC key cancels the error display and starts the operation with already registered set value.
• TEDS signal wires are incorrectly connected.
Refer to "8-2 CONNECTION" and check the connected state.
• Sensor connection is not appropriate.
Connect 1 to 4 sensors.
• Length of the cable too long
If the cable length is exceeding 30 m, the TEDS information may not correctly operate.
Take special care when using an extension cable to the sensor.
• Sensors not compatible with the TEDS are connected.
Cancel the automatic sensor registration function before use.
EID<CR>
None
ok<CR><LF>
When error occurs, E-03<CR><LF>