KYOWA WGA-670B-0 Instruction Manual page 43

Instrumentation amplifier
Table of Contents

Advertisement

(4) [MOD]
Contents:
Format:
Parameter:
Description & others:
Return value format:
a) Mode 00 return value format
Mode 00 repeatedly outputs the indicated value and status in order.
D1±xxxxxx<CR><LF>
Sabcdefgh<CR><LF>
• The return value ±xxxxxx is an indicated value with a decimal point.
If the indicated value has no decimal point, it is filled with '0' from the head.
• 'abcdefgh' describes status with '1' or '0.'
Numeric '1' is outputted to describe the following state.
a: Comparator H is [ON].
b: Comparator L is [ON].
c: Spare
d: Spare
e: During measurement
f: Detecting peak
g: During hold
h: Various errors occurred
• To stop the output, set the mode to Mode 10.
b) Mode 02 return value format
Mode 02 outputs the indicated value and status only once at hold.
D1±xxxxxx<CR><LF>
Sabcdefgh<CR><LF>
• Contents of the indicated value and status are as same as those of the Mode 00.
• Assume that it is operated in the Mode 00. If the Mode 02 is specified, the following message is sent
back.
ng<CR><LF>
c) Mode 10 return value format
Mode 10 operates the WGA-670B according to contents of the control commands.
ok<CR><LF>
Sets transmission mode
MODpp<CR>
pp (Transmission mode)
• 00: Repeatedly outputs the indicated value and status.
• 02: Outputs the indicated value and status only once at hold
mode.
• 10: Operation by control command.
This command becomes disabled with the power OFF. Power ON, and it is in
the Mode 10 operation mode.
38

Advertisement

Table of Contents
loading

This manual is also suitable for:

Wga-670bWga-670b-1Wga-670b-7

Table of Contents