10. ERROR CODE LIST

Error Display
0FL1
0FL2
0FL3
0FL4
E-00
E-01
E-02
E-03
Contents
Measured value is exceeding the
measuring range to the minus side.
Measured value is exceeding the
measuring range to the plus side.
Indicated value is below -19999.
Indicated value is above +19999.
Sensitivity
registering
exceeding the allowable range.
Memory (RAM) failure
Memory (EEP ROM) failure
TEDS
automatic
registration failure
Decrease the input signal.
Use the WGA-670B within
measuring range of ±3.2 mV/V.
Decrease the input signal or reset
the sensor registration.
value
is
Use
indication less than 10000 per
0.1 mV/V.
Failure
Contact
representatives for repair.
sensitivity
See item (4) in "8. TEDS
OPTION"
display is shown.
54
Countermeasures
the
WGA-670B
with
KYOWA
or
when
this
error
our