Servicing Repair And Disposal Information; Servicing - Lauda Proline RP 845 C Operating Instructions Manual

Table of Contents

Advertisement

9.3

Servicing repair and disposal information

9.3.1 Servicing

LAUDA Thermostats largely require no service. If the heat transfer liquid becomes contaminated, it
should be replaced ( Section 6.2).
102
 Withdraw the mains plug before all service and repair work.
 Only specialists must carry out repairs in the control section.
 Keep to service intervals ( 9.3.2). If servicing does not occur at the stated
intervals, then the manufacturer can no longer guarantee the safe operation of
the thermostatic circulator.
 At the back of the Proline head a main fuse switch
interrupts the mains connection when an overload occurs. It is then in the "o"
position and can be set in the "–" position again.
 If the fuse trips again, Service must locate the cause.
 Additionally, a safety fuse, which protects the low voltages, is situated on the
mains board. If a fuse fails ( mains lamp does not light) only replace with a
fuse with the specified data (one x T (= slow-blow) 10 A, size 5 x 20 Fuse is
located in the unit as shown below).