Servicing Repair And Disposal Informations; Servicing - Lauda RP 845 Operating Instructions Manual

Low-temperature thermostats with smartcool system
Table of Contents

Advertisement

9.3
Servicing repair and disposal information

9.3.1 Servicing

LAUDA Thermostats largely require no service. If the heat carrier liquid becomes contaminated, it
should be replaced
116
− Withdraw the mains plug before all service and repair work.
− Only specialists must carry out repairs in the control section.
Section 6.2.
− At the back of the Proline head a main fuse switch
interrupts the mains connection when an overload occurs. It is then in the "o"
position and can be set in the "–" position again.
− If the fuse trips again, Service must locate the cause.
− Additionally, a safety fuse, which protects the low voltages, is situated on the
mains board. If a fuse fails (
fuse with the specified data (one x T (= slow-blow) 10 A, size 5 x 20
located in the unit as shown below).
The SmartCool System refrigerating machine operates
largely without servicing. So that the full cooling power is
available, the heat exchanger (condenser) should cleaned of
dust at intervals of one month or longer depending on the
operating period and dust level in the ambient air. To do this,
open the front grille, brush off the condenser and, where
necessary, blow over with compressed air.
Extreme contamination is detected by the Proline SelfCheck
Assistant, which then issues a warning.
Maintenance
Proline Low-temperature Thermostats
mains lamp does not light) only replace with a
is located which
Fuse is
UL 533