AGFA DRY 3.4 Reference Manual page 43

Hide thumbs Also See for DRY 3.4:
Table of Contents

Advertisement

4. The 'Image quality' screen for the selected modality is displayed:
5. Press the Confirm key to select 'Look-up table'.
The 'Look-up table' screen for the selected modality is displayed:
6. Press the Up/Down arrow keys to select the desired lookup table.
7. Press the Confirm key to store the data, or the Escape key to quit the
procedure without any changes.
In case Kanamori-like is selected, change the Kanamori-value.
An additional screen for the Kanamori-like value is displayed:
All Kanamori-like curves are selectable starting from 75 until 220 in steps
of 1 unit. The activated setting is marked with an *.
Related Links
DRY 3.4 | Advanced Operation (key-operator mode) | 43
Note:
In case a number of settings for a modality were set via a
remote computer, you will not be able to change the image
quality information for this modality manually. If this is the
case, the following screen will appear:
Note: In case the image quality parameters need to be
changed for this modality, please use the DICOM
Configuration tool, in the Specialist or Service mode
(specialist mode available only via remote network PC). You
can also delete this modality and create a new modality with
the same name. For this new modality new parameters can
be assigned via the local keypad.
2604A EN 20180322 1454