nebulizer to free aspirate. Tighten the tension adjustment until the flow just stops then tighten by
one turn. Turn on the pump and, if necessary, increase the tension until a smooth flow is
produced. Do not over-tighten the pump clamps as it will result in excessive wear and tear of the
pump tubing and require replacement tubing at more frequent intervals.

7.9 Torch Alignment

For maximum sensitivity and optimum results it is advisable to check the alignment of the torch
whenever it has been removed and replaced in the instrument or if the torch body or centre tube
has been replaced.
Click on Qtegra to open the program. In the Dashboard Page, click Torch Alignment to start
the procedure. Introduce a "loaded Blank" sample (2 ppm Zn solution). And ensure you leave
enough time for the sample to enter the plasma before clicking ok. Aspirate until the process is
finished. The result of the test is displayed in the bottom (right side) of the page.

7.10 Reporting results

Reporting can be found in the Query tab of acquired LabBooks. Reports are global and not
saved per LabBook. The latest report style will be available for any LabBooks that are opened
subsequently (where they can also be modified).
To change general settings, click on the Report Options button and change the required
parameters:
Page settings – Paper format, Portrait or Landscape, Font size...