iCAP Customer Familiarisation Manual
Remove the Stator and the separator, noting their orientation:
Carefully remove the rotor, from the valve:
Using a clean cloth, gently clean the channels, ports and surfaces of the stator and rotor. If
necessary use low-pressure, canned air to blow the channels and ports free of any remaining
debris.
This document is part numbered 849940090071
Version 1.02
Page 53 of 72