iCAP Customer Familiarisation Manual
Note that any time the 6-port valve body is removed from its actuator; the valve will require
retraining (re-initialization). { XE "valve:retraining" }
To replace or re-orientate the 6-port valve:
Remove all of the tube connectors to the 6 port valve:
Using the provided hex key (9/64), loosen the hex screw on the locking collar which secures the
base of the valve to the instrument:
Firmly but carefully remove the valve head:
Insert the new valve, or reinsert the existing valve, at the desired angle. Rotate it so that the
nebulizer port will be as close as possible to the nebulizer.
This document is part numbered 849940090071
Version 1.02
Page 57 of 72