Centre Tube Insertion Into Torch Holder; Torch Holder Insertion Into The Torch Box; Positioning Of The Spray Chamber Drain - Thermo Scientific iCAP 7000 Series Manual

Icp-oes spectrometer
Hide thumbs Also See for iCAP 7000 Series:
Table of Contents

Advertisement

iCAP Customer Familiarisation Manual

5.4 Centre tube insertion into torch holder

Insert the centre tube assembly into the metal torch holder.
Screw the centre tube holder assembly in a clockwise direction into the metal torch holder until
the o-ring is compressed. Do not over tighten as this will reduce the lifetime of the O-ring seal.
When fitted the center tube should be 1-3mm lower than the intermediate tube as shown above.

5.5 Torch holder insertion into the torch box

Insert the torch holder into the
torch box and turn the metal torch holder clockwise until the red orientation lock self locates in the
torch box casting.

5.6 Positioning of the spray chamber drain

Insert the white plastic tubing connector and wide bore tubing (0.79mm inner diameter) into the
spray chamber drain tube. The drain and spray chamber should be positioned so that no pulsing
This document is part numbered 849940090071
Version 1.02
Page 11 of 72