iCAP 7000 Series ICP-OES Spectrometer

10 Performance Checks

10.1 Torch Alignment

Before running performance checks a torch alignment must be completed using the iCAP 7000
Set-Up Solution (Loaded Blank - 2ppm Zn, 0.01% Methanol).
Note: The iCAP test solutions are included with a new instrument and in the optional
Installation and Maintenance Kit.
NOTE: Torch alignment should be completed for any new torch or when the torch position is
disturbed.

10.2 Setting up Analyses

Note: An iCAP 7600 performance test must be completed with the valve deselected (effectively
the same as an iCAP 7400) The valve is tested separately.
Click on the Qtegra short cut to open the program.
On the Homepage page, select LabBooks.
Name a new LabBook "Service Installation Test".
Page 26 of 36
Torch alignment complete
Installation Details
Version 4301 460 00632