Sensor High trim by PC

Sensor High trim by P C is available us ing dedicated Software. ABB can provide the references and t he
Software.

Sensor Static trim

The purpose of this procedure is to o ptimize transmitter performance by reducing the effect of static line
pressure in these appli cations. 2600T Pressure transmitter Series does not normally require this procedure
because optimization based on facto ry characterization occurs in the sen sor. In any case this trim ca n be
performed if required.

Static trim by HHT

Sensor Static trim by Ha nd Held T erminal is avai lable. Please refer to the device man ual for the ri ght
operations.

Static trim by PC

Sensor Static trim by PC i s available using dedicated Software. ABB can prov ide the references and the
Software.

Calibration

Calibration is an operatio n by which t he LRV and URV are adjusted. It sets the 4 and 20 mA points at
required pressures.
See Basic Configuration Function, Set the LRV and Set the URV for the procedures.

Output trimming (Digital to Analog trim)

The Analog Output Trim allows the adjustment of the transmitter's current output at the 4 and 20 mA points
to match the plant standards. This command adjusts the digital to analog signal conversion.

Output trimming by Digital LCD Integral Display

To set the 4 mA output current
Connect the pre ssure transmitter as defined in the Trimming and Calibration Set Up. Using a
milliamperometer read the current output. Navigate the Digital LCD Integral Display menu, select
Output Trim
>
Set 4 mA
with the up a nd down keys and confirm with
same reading of the milliamperometer and confirm with
Note: the Maximum and Minimum values are indicated on the Display as reference.
Digital LCD Integral Display Keys sequence to set the 4 mA output current: 6 – 2 – 1 – Edit
To set the 20 mA output current
Connect the pre ssure transmitter as defined in the Trimming and Calibration Set Up. Using a
milliamperometer read the current at t he output. Navigate the Digital LCD Integral Display menu, select
Calibrate
>
Output Trim
desired value with the up and down keys and confirm with
have the same reading of the milliamperometer and confirm with
Note: the Maximum and Minimum values are indicated on the Display as reference.
Digital LCD Integral Display Keys sequence to set the 20 mA output current: 6 – 2 – 2 – Edit

Output trimming by CoMeter

Output trimming by CoMeter is available with the 695 Field Indicator. Please refer to the device manual for
the appropriate operations.

Output trimming by HHT

Output trimming by Hand Held Terminal is available. Please refer to the device manual fo r the appropriate
operations.
and the value will be available. To modify select
Next.
>
Set 20 mA
and the valu e will be avai lable. To modify select
Edit
Repeat the operation for the seven di gits to have t he
OK.
Next.
Repeat the operation for the seven digits to
OK.
364Ax-Gx Instruction Manual
Calibrate
then scroll th e desired value
Edit
then scroll the
precison
>
precison