Configuration

Default configuration

The data default configuration at the product delivery is the following:
Damping:
Units:
LRV:
URV:
Transfer function:
Hardware setting:
Low saturation current:
High saturation current: 20.5 mA
Low fail current:
High fail current:
LCD Display:

Easy Setup

To facilitate commissioning the 364 Display has the capability to run a sequence of predefined setup steps
with the main configuration settings.
When you start the Easy Setup you must continue up to the end to exit.
To start the Easy Setup:
Navigate the Digital LCD Integral Display menu, select
Digital LCD Integral Display Keys sequence to start the Easy Setup: 1
Define the language
If the desired language is different from the default o ne select
the following table) scrolling with the up and down keys and confirming with OK. Select
English
German
Define the Engineering Unit
If the desired Engineering Unit is different from the i ndicated, select
the following table) scrolling up and down keys and confirm with OK. Select
mbar
millibar
2
g/cm
grams per square centimeter
2
Kg/cm
kilo grams per square centimeter
Pa
pascal
kPa
kilopascal
torr
torr
atm
atmosphere
MPa
Megapascal
inH2O°C
inches of water at 4 degrees Celsius
Define the Lower Range value (LRV)
If the desired LRV is different from the default one select Edit, then choose the desired value for the digit in
reverse color with the up and down keys and confirm with
and confirm with
OK.
Select
Note: the minimum and the maximum allowed values are indicated on the display.
Define the Upper Range Value (URV)
If the desired URV is different from the i ndicated select Edit, scroll the desired value for the digit in reverse
color with th e up and down keys an d confirm with
confirm with
OK.
Select
Note: the minimum and the maximum allowed values are indicated on the display.
Run the Zero Scaling
In case a Zero Scaling is required a pply the pressure for the Z ero and select OK. Wait for th e auto-set end
(the bargraph will indicate the working progress). Select
Define the Damping
If the desired Damping is different from the indicated value change it with the up and down keys and confirm
with
OK.
1 second
kPa
0 kPa
equal to URL
linear
Write protect mode in OFF, Failsafe direction High
3.8 mA
3.7 mA
22 mA
Set to display the PV value.
Spanish
French
Next
to continue.
Next
to continue.
Easy
Setup.
Edit
then choose the desired language (see
Edit
mmH2O°C
millimeter of water at 4 degrees Celsius
inH2O°F
inches of water at 68 degrees Fahrenheit (20°C)
inHg°C
inches of mercury at 0 degrees Celsius
ftH2O°F
feet of water at 68 degrees Fahrenheit (20°C)
mmH2O°F
millimeter of water at 68 degrees Fahrenheit (20°C)
mmHg°C
millimeter of mercury at 0 degrees Celsius
psi
Pounds per square inch
bar
Bars
Next.
Repeat the operation for the seven digits
Next.
Repeat the operation for the seven digits a nd
Next
to continue.
364Ax-Gx Instruction Manual
to continue.
Next
Italian
then choose the desired unit (see
Next
to continue.
23