EMS AIR-FLOW S1 Operation Instructions Manual page 85

Hide thumbs Also See for AIR-FLOW S1:
Table of Contents

Advertisement

Air flow from the handpiece without
pressing the footpedal
Air and / or powder is leaking from
the threads of the cap of powder
chamber
Water leakage between handpiece
cord and handpiece
Leakage on the unit water supply
FB-303_4_ed_2014_07_AFS1.indd 69
• Send the unit to an approved EMS repair center
• Check the seal and the cleanness on the thread of the powder chamber and of the cap
• Replace the seal if necessary
• Clean the cap and the powder chamber
• Change the cap
• Send the unit to an approved EMS repair center
• Check the O-ring on the handpiece cord
• Send the handpiece cord to an approved EMS repair center
• Check the water supply connection
• Change the water supply hose with connector
• Send the unit to an approved EMS repair center
69
9/9/2014 11:23:49 AM

Advertisement

Table of Contents
loading

Table of Contents