82
6. When the Raw Input stops decreasing, press the Enter key to save the value (or press
to cancel).
7. Press the Set Span selector button. A highlighted edit box opens:
B
Calibrate dO2 0
i
o
F
l
o
1
1
0
8. Using the touchpad, key 100 for the new Span value into the highlighted edit box. Do
not press Enter yet.
9. Establish a reference 100% level of dissolved oxygen in the media: sparge with air at
the appropriate flow rate for your application.
10. When the Raw Input stops increasing, press the Enter key to save the 100% dO2
reference value.
12. After the values have been set, press the
6.7

Powering the Jacket Water Heater

If you have a non-jacketed vessel, skip this section. After sterilization, you should have
installed the vessel on the jacket water heater and connected the heater to the Power
Controller (see Section 6.4). Never set the temperature control loop to the desired
working temperature (refer to Section 7.4.1) until all water connections have been made,
the jacket is filled with water, the vessel is filled with media, and the RTD has been
installed in the thermowell and connected to the Power Controller.
Never plug the jacket water heater into the controller before the vessel is filled. When the
heater is connected properly and the PCU is turned on, the jacket water stirrer should spin
freely.
WARNING!
When the jacket water heater is plugged in, do not touch the top surface.
It is very hot.
BioFlo 110 Modular Fermentor M1273-0054
Zero Std: 0.00
Raw Input: 1022
Set Zero
Set Span
Reset
button to return to the Main Screen.
User's Guide