Advanced Control
0
8. Press the Agit selector button. The Setup screen returns, showing the new
cascade parameter:
B
dO2 73.1 %
i
o
F
l
o
1
1
Advanced Control
0
9. If you wish to change the high and low limits for Agitation speed, use the Agit
Hi and Agit Lo selector buttons to open an edit box for each, use the touchpad
to input a value, and press the Enter button to save the new value to memory.
The Hi Limit may not exceed 300 and the Lo Limit may not go below 25.
10. Press
Screen.
To cascade to 4 Gas: follow steps 1-4 above, but select 4 Gas instead of
Agitation.
To cascade to Demand: follow steps 1-4 above, but select Demand instead of
Agitation.
BioFlo 110 Modular Fermentor M1273-0054
Cascade To: None
Agit Hi Limit: 300
Agit Lo Limit:
25
P-Gain: 0.20
I-Gain: 1.00
Cascade To: Agit
Agit Hi Limit: 300
Agit Lo Limit:
25
P-Gain: 0.20
I-Gain: 1.00
to return to the dO2 Details screen or
Agit
4 Gas
Demand
None
Defaults
Cascade
Agit Hi
Agit Lo
P Gain
I Gain
to return to the Main
User's Guide