2. In the Gas loop screen that opens, set the O2 Mix in Auto Mode, as reflected in
the resulting screen:
B
Gas 0.0
i
o
F
l
o
1
1
0
3. The default gas flow setpoint is 20. If you wish to change it, select Setpoint,
then type in the desired value (from 0 to 20 SLPM—the system will
automatically limit the gas setpoint to 20).
4. Select Set O2% to set the percentage of oxygen.
NOTE:
After you set the O2%, the PCU will provide the complementary percentage of
air. For example, if you set O2% at 25, the PCU will provide 75% air.
BioFlo 110 Modular Fermentor M1273-0054
Value
Setpoint
23.4
30.0 °C
134
350 rpm
0.0
0.0 %
0.0
0.0 %
0.0
0.0 %
6.68
7.00 pH
77.1
0.0 %
0.3 20.0 SLPM Auto
Control: Off
Setpoint: 0.0
Output: 0.0
O2 Mix: Auto
Output: 0.0
Control
Off
Auto
Manual
Manual
Manual
Auto
Auto
Press here
SLPM
Control
Setpoint
O2 Mix
Set O2%