NEW BRUNSWICK SCIENTIFIC BioFlo 110 Manual To Operations page 39

Table of Contents

Advertisement

26
1. Wear protective gloves to protect yourself in case of accidental breakage.
2. Lightly coat the pH probe with glycerol.
CAUTION!
Do not install the pH port adaptor in the headplate before inserting the
probe. Follow the steps below to fit the pH port adapter onto the probe
first, then insert the probe and adapter into the headplate.
Figure 15: pH Probe with Port Adapter (exploded)
BOTTOM FERRULE
With reference to Figure 15 above:
3. Gently slide the top portion of the knurled port adapter (part of the probe kit)
onto the probe.
4. Slide the two white ferrules onto the probe, the narrower one on top of the
deeper, cup-shaped one.
5. Gently slide the bottom portion of the port adapter onto the probe, taking care
to orient the longer threaded section toward the top of the probe.
6. Remove the two O-rings installed in the pH port; first slide the white Teflon O-
ring onto the probe, then follow with the black 12mm port adapter O-ring.
7. Do not yet close up all the elements of the port adapter.
8. Gently insert the probe into the appropriate port (see Figures 11, 12 & 13 for
reference), allowing the O-rings to seat fully into the port.
NOTE:
The fit may be snug. Gently rotate the probe as you press it into the port to
avoid breakage.
BioFlo 110 Modular Fermentor M1273-0054
CAP
pH PROBE ADAPTER
(top portion)
TOP FERRULE
pH PROBE ADAPTER
(bottom portion)
TEFLON O-RING (white)
PORT O-RING (black)
User's Guide