Conversely, when the actual dO2 value drops below the setpoint, the cascade
system acts to bring it back up.
• 4 Gas controls dissolved oxygen by automatically adjusting the feed of all
gases introduced to the culture. (This is not available without the Gas Mix
Controller.)
• Demand controls dissolved oxygen by automatically adjusting the feed of
carbon dioxide, nitrogen and oxygen. (This is not available without the Gas
Mix Controller.)
• The default setting is Cascade To: None, which means that dissolved oxygen
will be controlled only by the addition of oxygen when the actual dO2 level
drops below setpoint.
To cascade dissolved oxygen to Agitation:
6. In the dO2 Details screen, press the Setup selector button. The Setup screen
opens:
B
dO2 73.1 %
i
o
F
l
o
1
1
Advanced Control
0
7. Press the Cascade selector button. The Cascade screen opens:
New Brunswick Scientific
Cascade To: None
Agit Hi Limit: 300
Agit Lo Limit:
25
P-Gain: 0.20
I-Gain: 1.00
103
Defaults
Cascade
Agit Hi
Agit Lo
P Gain
I Gain