NEW BRUNSWICK SCIENTIFIC BioFlo 110 Manual To Operations page 124

Table of Contents

Advertisement

B
Pump C 0.0 %
i
Control: Off
o
Output: 0
F
100
l
o
1
1
0
0
2. Press the Antifoam selector button. The pump Details screen screen returns;
after a short delay, the control mode will reflect your choice. Note that a new
parameter, Probe Sensitivity, also appears:
B
Pump C 0.0 %
i
Control: Foam
o
Probe Sens.: 75.0
F
100
l
o
1
1
0
0
When the probe
is activated, CV
3. Probe sensitivity reflects the density of foam necessary to activate the antifoam
is current
pump. To set a sensitivity value (we recommend starting with 75.0), press the
percent output
Sensitivity selector button, input the desired value using the touchpad, and
(either 0 .0 or
press the Enter key.
the setpoint).
4. Press the
5. Follow the same procedure to set the Wet On pump and the probe sensitivity
value, and the Wet Off pump and probe sensitivity.
New Brunswick Scientific
CV
SP
0.0
0.0
CV
SP
0.0
75.0
button to return to the Main Screen.
Off
Manual
Antifoam
Wet On
Wet Off
Control
Setpoint
Disp Hi
Disp Lo
Sensitivity
User's Guide
111