3.1
Powering Up
To initially start the instrument, push the
NOTE:
The LED display will be flashing, indicating that the unit has been off (or
that the power to the unit has been interrupted). To stop the flashing
display, press any key on the master control panel keypad.
3.2
Temperature
SELECT
Press the
current chamber temperature. The temperature can be set from 4.0°C to 60.0°C, using
the
or the
The temperature alarms, both audible and visible (flashing light), are activated if the
temperature is more than 1.0°C above or below the temperature setpoint. The alarm
will automatically deactivate as the unit achieves the set temperature.
3.3
Relative Humidity
SELECT
Press the
current relative humidity inside the chamber. The relative humidity can be set from 0
to 90% RH. Increasing or decreasing the setpoint is accomplished with the
key.
The relative humidity alarms, both audible and visible (flashing light), are activated if
the relative humidity is more than 5% RH above or below the setpoint. The alarm will
automatically deactivate as the chamber achieves the set relative humidity.
3.4
Alternate Temperature Setpoint
If a lamp manifold option is installed, the Innova 4900 allows a secondary or alternate
temperature setpoint to be entered:
M1197-0053
Innova 4900 Environmental Chamber
°C
key until the
function indicator lights. The display shows the
key.
%RH
key until the
function indicator lights. The display shows the
3
O
3
O
P
E
P
E
ON/OFF
switch to the
25
R
A
T
I
O
N
R
A
T
I
O
N
ON
position.
or the
User's Guide