Filling The Water Jacket; Install Baffle (14.0L Fermentation Vessels Only) - NEW BRUNSWICK SCIENTIFIC BioFlo CelliGen 115 Manual To Operations

Benchtop fermentor & bioreactor
Table of Contents

Advertisement

Figure 10: Water Jacket Guard Installation (top view)
Rubber Gasket
M1155-9902 (1.3L & 3.0L)
M1227-9903 (7.5L & 14.0L)
5. With the gradations marked on the glass facing front (toward the user),
slide the vessel into the bottom clamping ring, until it rests securely against
the base plate. Make sure the water inlet tube stands free (not kinked)
inside the water jacket.
6. Finger tighten the six knurled thumb screws, to securely attach the
clamping ring to the base plate. This seals the water jacket.
4.8.3
To fill the water jacket:
1. After the tubing and water supply are connected, make sure the solenoid
valve cable and the RTD cable are plugged into the Power Controller.
2. Set the temperature control mode to Off.
3. Check that the temperature reading is higher than 5ºC.
4. Allow water to enter the piping system; it will stop at the solenoid valve.
5. Set the temperature loop control mode to Auto.
6. Enter a temperature setpoint (
value (
opening the solenoid valve, filling the jacket with water.
4.8.4
For installation of the 1.3L, 3.0L & 7.5L vessel baffle, see Section 4.8.21.
If you are using a 14.0L vessel, install the baffle assembly inside the glass
vessel:
New Brunswick Scientific

Filling the Water Jacket

). The controller will respond to the call for cooling by
CV

Install Baffle (14.0L Fermentation Vessels ONLY)

) that is at least 12ºC below the current
SP
29
Bottom
Vessel
Flange
User's Guide