NOTE:
The water solenoid valve's brass body has two hose barbs, identified as "1"
and "2" by numbers stamped on the body, just above the barb. Hose barb 1
is the water inlet. Hose barb 2 is the water outlet. The valve body swivels to
facilitate access.
8. Connect silicone tubing to the top outlet port of the exhaust condenser. Allow
MOUNTING
BRACKET
HOSE BARB 1:
connect to water
source, downstream
of pressure regulator
VALVE BODY
STERILE
FILTER
WATER INLET
New Brunswick Scientific
enough tubing to run to the water drain. Secure the tubing with a plastic tie.
Figure 17: Water Solenoid Valve
Figure 18a: Exhaust Condenser
(1.3L, 3.0L & 7.5L Vessels)
WATER
OUTLET
EXHAUST PORT
VALVE SPLITTER
HOSE BARB 2:
connect to cooling
coil or water jacket
inlet
SET SCREW
in port adapter
PLUG: connect
to rear panel of
Power Controller
(see Figure 2)
HEADPLATE
User's Guide
29