NEW BRUNSWICK SCIENTIFIC BioFlo 110 Manual To Operations page 61

Table of Contents

Advertisement

48
5. Attach the exhaust filter (respecting the direction of flow if stamped on the
filter) to the condenser outlet. Secure the filter with a plastic tie.
6. With reference to Figures 30a and 30b, connect a length of silicone tubing
between the lower (cooling water inlet) port of the exhaust condenser and the
T-connector (upstream of the water solenoid valve). Secure at both ends with a
plastic tie.
7. Connect the T-connector inlet (see Figure 30b) to hose barb 2 (see NOTE and
Figure 29 below) on the water solenoid valve. The valve is on the side of the
Power Controller. Secure with a plastic tie.
NOTE:
The water solenoid valve's brass body has two hose barbs, identified as "1"
and "2" by numbers stamped on the body, just above the barb. Hose barb 1
is the water inlet. Hose barb 2 is the water outlet. The valve body swivels to
facilitate access.
8. Connect silicone tubing to the top (cooling water outlet) port of the exhaust
condenser. Allow enough tubing to run to the water drain. Secure the tubing
with a plastic tie.
MOUNTING
BRACKET
HOSE BARB 1:
connect to water
source, downstream
of pressure regulator
VALVE BODY
BioFlo 110 Modular Fermentor M1273-0054
Figure 29: Water Solenoid Valve
VALVE SPLITTER
HOSE BARB 2:
connect to cooling
coil or water jacket
inlet
PLUG: connect
to rear panel of
Power Controller
(see Figure 2)