NEW BRUNSWICK SCIENTIFIC BioFlo 110 Manual To Operations page 138

Table of Contents

Advertisement

Weight
Dimensions
Agitation
Temperature
Electrical
Output Connectors
Inputs
Power Switch
Pilot Light
Other
Weight
Dimensions
Inputs
Control Options
Power
New Brunswick Scientific
BioFlo 110 Power Controller
Height
Depth*
Width
*allow 3 inches (7.6 cm) more, front and rear, for cable access
Drive
Speed Range
Impellers
Standard Rushton; optional pitched blade
Range
5°C above ambient temperature to 70°C
System
20-40°C ± 0.1°C (± 0.5°C at other temperatures)
Accuracy
Control Stability
0.1°C
Sensor
One 100 platinum RTD, input located on front of console
Control
P-I control algorithm
SSR PWM
For heat blanket and for cooling solenoid valve
Outputs
115/230 VAC, 50-60 Hz, 1500 VA
Located on front.
Located on front. Illuminates when power is on.
BioFlo 110 dO2/pH Controller
Height
Depth*
Width
*allow 3 inches (7.6 cm) more, front and rear, for cable access
pH Probe
NBS-supplied cable
Dissolved
Polarographic electrode with NBS-supplied cable.
Oxygen Probe
See PCU Specification
12 VDC (supplied by PCU)
16 lb. (7.2 kg)
8 9/16 in. (21.8 cm)
12 in. (30.5 cm)
12 in. (30.5 cm)
Lift-off electric motor.
Self-aligning motor and vessel shafts.
Direct drive with surface seal for fermentation and
cell culture.
Magnetic drive for cell culture.
Fermentation: 50-1200 rpm
Cell culture:
30-300 rpm
Five controlled (PWM) IEC outlets for pumps
Vessel Heater
Vessel Cooling Solenoid Valve
One switched mains outlet for PCU
Agitation motor
Tachometer (joint with agitation output)
RTD (temperature probe)
Hangers for 2 Rotameters.
Hanger for cooling water valve.
Cable/tubing organizers.
5 lb. (2.3 kg)
2 5/8 in. (6.7 cm)
12 in. (30.5 cm)
12 in. (30.5 cm)
125
User's Guide

Hide quick links:

Advertisement

Table of Contents

Troubleshooting

loading