Calibration Certificate ID
072731-239-041621-ACC-USI
Measurement Uncertainty of the Weighing Instrument in Use
Stated is the expanded uncertainty with k=2 in use. The formula shall be used for the estimation of the uncertainty under consideration of
the errors of indication. The value R represents the net load indication in the unit of measure of the device.
Temperature coefficient for the evaluation of the measurement uncertainty in use:
Temperature range on site for the evaluation of the measurement uncertainty in use:
Linearization of Uncertainty Equation
Range
d
1
0.02 lb
To optimize the stability of the linearization, besides of the zero load only increasing measurement points with a test load of 5% of the
measurement range or larger are taken for the calculation of the linear equation.
Absolute and Relative Measurement Uncertainty in Use for Various Net Indications (Examples)
Net Indication
1.00 lb
10.00 lb
100.00 lb
500.00 lb
1000.00 lb
Software Version: 1.23.0.77
Report Version: 2.13.4
Max
1000 lb
N/A
N/A
N/A
N/A
N/A
As Found
This is an original document and may not be partially reproduced without the
written permission of the issuing calibration laboratory.
As Found
N/A
As Found
N/A
N/A
N/A
N/A
N/A
As Left
© METTLER TOLEDO
10.0 · 10
-6
/ K
10 K
As Left
U
= 0.023 lb + 0.0000986 lb/lb · R
1
As Left
0.023 lb
2.3%
0.024 lb
0.24%
0.033 lb
0.033%
0.072 lb
0.014%
0.12 lb
0.012%
Page 5 of 5