Easy Calibration (Adjustment) - BRAND Transferpette electronic Operating Manual

Motor diven single channel pipette
Hide thumbs Also See for Transferpette electronic:
Table of Contents

Advertisement

Easy Calibration
The calibration mode ‚CAL'
Adjustment
The instrument should be set to either the nominal volume (for example
200 µl for a 200 µl pipette) or a specific test volume, in the standard
pipetting mode (PIP). See page 42 and 46 for procedures.
E.g., volume according to testing of volume 201.3 µl.
What to do
1.
Bring up the CAL
mode
2.
Activate the CAL
mode
3.
Confirm CAL mode
4.
Set the volume
5.
Confirm volume
58
How to do it
Press and hold the MENU
key (> 3 sec) until CAL
appears. The display reads
‚off'. ‚CAL' blinks.
Press one of the arrow
buttons to activate the CAL
mode. The display changes
from ‚off' to ‚on'. ‚CAL' con-
tinues to blink.
Press the ENTER key. The
display now shows the set
pipetting volume.
‚CAL' blinks.
Use the arrow keys (+/-) to
set the volume, which was
previously determined and
tested. ‚CAL' blinks.
Press the ENTER key.
The display shows the tested
and corrected volume. The
CAL symbol is continously
displayed to confirm that an
adjustment has been made.
Adjustment of the Transferpette
Keys to press
1x
>3s
1x
1x
electronic
®
Display readout