Adjustment - Easy Calibration - BRAND Transferpette S Operating Manual

Micropipettes
Hide thumbs Also See for Transferpette S:
Table of Contents

Advertisement

7 Adjustment – Easy Calibration
7 Adjustment – Easy Calibration
The instrument is permanently calibrated for aqueous solutions. If it is de-
termined that the pipette is operating inaccurately or to adjust the instru-
ment to work with solutions of varying density and viscosity or with spe-
cially-shaped pipette tips, it can be calibrated using the Easy Calibration
Technique.
b.
d.
Operating manual
50
a.
1
b.
2
3
c.
d.
Perform a volume check and deter-
mine the actual value; see Checking
the volume.
Remove label window (1) and label-
ing film (2): Gently move the clamp
and lift it off.
Using a paper clip or an unused
pipette tip, remove the protective
film (3) (the protective film can be
discarded).
Slide the red adjustment slider back
completely, lift the volume-setting
wheel (decoupling) and release the
adjustment slider.
Transferpette® S 997207