Easy Calibration (Adjustment) - BRAND Transferpette -8 Operating Manual

Table of Contents

Advertisement

Available languages
  • EN

Available languages

  • ENGLISH, page 39
Easy Calibration
The calibration mode ‚CAL'
Adjustment
The instrument should be set to either the nominal volume (for example
100 µl for a 100 µl pipette) or a specific test volume, in the standard
pipetting mode (PIP). See page 44 and 48 for procedures.
E.g., volume according to testing of volume 101,3 µl.
What to do
1.
Bring up the CAL
mode
2.
Activate the CAL
mode
3.
Confirm CAL mode
4.
Set the volume
5.
Confirm volume
60
Adjustment of the Transferpette
How to do it
Press and hold the MENU
key (> 3 sec) until CAL
appears. The display reads
‚off'. ‚CAL' blinks.
Press one of the arrow
buttons to activate the CAL
mode. The display changes
from ‚off' to ‚on'. ‚CAL' con-
tinues to blink.
Press the ENTER key. The
display now shows the set
pipetting volume.
‚CAL' blinks.
Use the arrow keys (+/-) to
set the volume, which was
previously determined and
tested. ‚CAL' blinks.
Press the ENTER key.
The display shows the tested
and corrected volume. The
CAL symbol is continously
displayed to confirm that an
adjustment has been made.
Keys to press
Display readout
1x
>3s
1x
1x
-8/-12 electronic
®

Advertisement

Table of Contents
loading

This manual is also suitable for:

Transferpette -12

Table of Contents