7 Adjustment – Easy Calibration
7 Adjustment – Easy Calibration
The instrument is permanently calibrated for aqueous solutions. If it is de-
termined that the pipette is operating inaccurately or to adjust the instru-
ment to work with solutions of varying density and viscosity or with spe-
cially-shaped pipette tips, it can be calibrated using the Easy Calibration
Technique.
b.
d.
Operating manual
56
a. Perform a volume check and deter-
1
2
b. Remove label window (1) and label-
3
c. Using a paper clip or an unused
d. Slide the red adjustment slider back
mine the actual value; see Checking
the volume.
ing film (2): Gently move the clamp
and lift it off.
pipette tip, remove the protective
film (3) (the protective film can be
discarded).
completely, lift the volume-setting
wheel (decoupling) and release the
adjustment slider.