Input
10.1.1
Measured variable
Level limit detection of change in capacitance between probe rod and container wall or ground tube,
depending on the level of a liquid.
10.1.2
Measuring range (valid for all FEI5x)
• Measuring frequency:
500 kHz
• Span:
∆C = 0 to 1600 pF
• Final capacitance:
C
= max. 1600 pF
E
• Adjustable initial capacitance:
C
= 0 to 500 pF (range 1 = factory setting)
A
C
= 0 to 1600 pF (range 2)
A
10.1.3
Input signal
Probe covered
=> high capacitance
Probe not covered => low capacitance
10.2
Output
10.2.1
Galvanic isolation
FEI52:
Between rod probe and power supply
FEI54:
Between rod probe, power supply and load
FEI53, FEI55, FEI57S:
See connected switching unit
10.2.2
Switch behavior
Binary or ∆s operation (pump control)
10.2.3
Switch-on behavior
When the power supply is switched on, the switching status of the outputs corresponds to the signal
on alarm.
After max. 3 s the correct switching status is achieved.