With Separate Housing - Endress+Hauser Liquicap M FTI51 Operating Instructions Manual

Hide thumbs Also See for Liquicap M FTI51:
Table of Contents

Advertisement

Installation
!
32
3.10

With separate housing

For information on ordering, refer to "Ordering information" => "Probe design"
E E X X
Zone 1
Zone 1
Note!
• The maximum connection length between the probe and the separate housing is 6 m (L4).
The desired length must be quoted when ordering a Liquicap M with a separate housing.
• If the connecting cable is to be shortened or guided through a wall, it has to be separated from
the process connection. For more information, refer to → Chap. 3.11.1.
• The cable has a bending radius of r ≥ 100 mm which may not be undershot!
3.10.1
Extension heights
Housing side: wall mounting
~61 mm
B
L4
L4
£ £
6 m
6 m
Rod length L1 max. 4 m
Rod length L1 max. 10 m
Housing side: pipe mounting
B
~75 mm
Liquicap M FMI51, FMI52
E E X X
Zone 0
Zone 0
L00–FMI5xxxx–14–00–06–xx–003
Sensor side
r
100 mm
D
r
100 mm
L00–FMI5xxxx–06–05–xx–xx–049
Endress+Hauser

Advertisement

Table of Contents
loading

This manual is also suitable for:

Liquicap m fti52

Table of Contents