Connecting The Electronic Insert Fei52 (Dc Pnp) - Endress+Hauser Liquicap M FTI51 Operating Instructions Manual

Hide thumbs Also See for Liquicap M FTI51:
Table of Contents

Advertisement

Wiring
42
4.6

Connecting the electronic insert FEI52 (DC PNP)

The three-wire DC connection should, wherever possible, be connected as follows:
• To programmable logic controllers (PLCs),
• To DI modules in accordance with EN 61131–2
A positive signal is present at the switch output of the electronic system (PNP).
Power supply
DC voltage: 10 V...55 V
Ripple: max. 1.7 V, 0...400 Hz
Current consumption: < 20 mA
Power consumption: max. 0.9 W
Reverse polarity protection: yes
Separation voltage: 3.7 kV
FEI52 overvoltage protection: overvoltage category III
Signal on alarm
Output signal on power failure or in the event of device failure: I
Connectable load
• Load switched via transistor and separate PNP connection, max. 55 V
• Load current max. 350 mA (cyclical overload and short-circuit protection)
• Residual current < 100 µA (with transistor blocked)
• Capacitive load max. 0.5 µF at 55 V, max. 1.0 µF at 24 V
• Residual voltage < 3 V (for transistor switched through)
Connect the FEI52 (DC PNP) as follows:
1.
Make the connection as shown in the graphic.
2.
Turn the cable gland until tight.
3.
Set the function switch (5) to position 1
(operation).
!
Note!
Do not switch on the supply voltage until you have
familiarized yourself with the device functions as
described in chapter 5, "Operation". This will ensure
that you do not accidentally trigger any processes by
switching on the supply voltage.
4.
Switch on the supply voltage.
* R = External load (I
350 mA, U
max.
F F
0.5A
0.5A
55 V DC
max
Liquicap M FMI51, FMI52
< 100 µA
R
FEI
FEI52
52
1 2
1 2
3
3
(+)
(+)
R R
– –
L+
L+ L–
L–
U – 10...55 V (DC)
U – 10...55 V
(DC)
... ...
L00–FTI5xxxx–04–05–xx–xx–007