MODBUS (RS485)
Parameters

MODBUS/TCP Parameters

Programming Site Data
If your DF868 does not include an option card for MODBUS
communications, you have completed programming the COMM
submenu. However, if you have installed a MODBUS option card, the
following additional prompts appear.
[F1]-[F4]
1. Press
to select the desired MODBUS Baud Rate from four
selections: 2400, 4800, 9600 and 19,200.
[F1]-[F3]
2. Press
to set the MODBUS Parity as none, odd or even.
[F1]
3. Press
to select one MODBUS Stop Bit, or
stop bits.
4. Use the numeric keys to enter a MODBUS Address (from 1 to 247)
[ENT].
and press
If your DF868 does not include an option card for MODBUS/TCP
communications, you have completed programming the COMM
submenu. However, if you have installed a MODBUS/TCP option
card, complete the following steps:
[F3]
1. Press
to select the appropriate MODBUS Baud Rate of 9600.
[F1]-[F3]
2. Press
to set the MODBUS Parity as none, odd or even.
[F1]
3. Press
to select one MODBUS Stop Bit, or
stop bits.
4. Use the numeric keys to enter a MODBUS Address (from 1 to 247)
[ENT].
and press
March 2008
[F2]
to select two
[F2]
to select two
1-55