Advertisement

Programming a Totalizer
Output
Programming Site Data
This type of output issues one pulse per selected volume of flow. The
meter produces a pulse each time the programmed amount of flow
passes through the pipe.
[F1]-[F4]
9. Press
to select the parameter to be totalized. See Table 1-
24 below for a description of the available options.
Table 1-24: Output Measurement Options
Option Bar Choice
=
[F1]
+TOTL
=
[F2]
-TOTL
=
[F3]
+MASS*
=
[F4]
-MASS*
→]
= +ENRG**
[
+ [F1]
→]
= -ENRG**
[
+ [F2]
*Available only if
Available only if
**
Note: The measurement units that appear in these prompts are those
GLOBL-SYSTM
selected in the
10.Enter a value from 50 to 4,000,000 μsec for the Minimum
of each totalizer pulse and press
Note: A complete pulse consists of equal amounts of
times. Choose a value that is compatible with the frequency
counter to be used.
11. Enter a value for the number of measurement Units represented by
each pulse and press
[F1]
[F4]
12.Press
-
to set up another totalizer/frequency output or press
[EXIT]
to select another slot for set up.
Description
Forward Totalized Volume Flow
Reverse Totalized Volume Flow
Forward Totalized Mass Flow
Reverse Totalized Mass Flow
Forward Energy Flow
Reverse Energy Flow
Mass Flow (static)?
Energy Option = YES
menu earlier in this chapter.
[ENT]
.
[ENT]
.
March 2008
=
YES
ON
Time
ON
OFF
and
1-51