GE DigitalFlow DF868 Programming Manual page 56

Panametrics multipurpose ultrasonic liquid flowmeter
Hide thumbs Also See for DigitalFlow DF868:
Table of Contents

Advertisement

March 2008
Slot X Analog Outputs (cont.)
1-48
DIAG
The
option provides access to a wide variety of diagnostics
parameters. See Chapter 3, Diagnostics, of the Service Manual for a
complete description of these options.
Note: The measurement units that appear in these prompts are those
GLOBL-SYSTM
selected in the
5. Enter a value for the Base (low end) of the chosen output range
[ENT]
and press
.
6. Enter a value for the Full (high) end of the chosen output range
[ENT]
and press
.
[F1]-[F4]
7. Use the
keys to select the desired Error Option, the way
the display responds during an error condition. Available options
are shown in Table 1-22 below.
Table 1-22: Error Response Options
Option
HOLD
Holds the last "good" reading.
LOW
Forces the outputs to the low setpoint.
HIGH
Forces the outputs to the high setpoint.
OTHER
Allows user to enter 4-20 mA error level in mA.
[F1]
[F4]
8. Press
-
to set up another output or press
another slot for set up.
For instructions on setting up additional inputs and/or outputs
proceed to the appropriate section. Otherwise, press
Global Program
return to the
menu earlier in this chapter.
Display Response
[EXIT]
[EXIT]
prompt.
Programming Site Data
to select
twice to