Assembly of Other Devices (Optional)

3.2
1
Attach a nosepiece.
Note on device combinations
The nosepiece cannot be used in combination with the microscope system that includes any of the following devices:
 C-FMAN/C-FMBN Focus Mount
 C-LEDS Hybrid LED Stand
To use two objectives, attach a nosepiece to the zooming body.
Attach a nosepiece before performing 3.1, step 6 "Install an objective."
The P-RN2 Nosepiece is intended for the SMZ800N Zooming Body or the SMZ1270 Zooming Body, and the P-RNI2
Intelligent Nosepiece is intended for the SMZ1270i Zooming Body. The procedure for attaching a nosepiece is the
same for both types.
(1)
First check that the nosepiece is in the stereoscopic
view state as shown in the figure.
Note that the nosepiece in the vertical view state as
shown in the figure cannot be attached.
(2)
Align the positioning pin for attaching the nosepiece
with the positioning groove at the bottom rear of the
zooming body so that the nosepiece fixing screw fits
the objective fixing screw at the bottom of the
zooming body. Insert the attachment tool along the
two grooves from the bottom of the nosepiece and
tighten the nosepiece fixing screw to securely attach
the nosepiece by turning the attachment tool.
Chapter 5
Assembly
Nosepiece in the stereoscopic view state
71
(The P-RN2 is used as an example.)
Nosepiece in the vertical view state
(Attachment not possible in this state)
Securely attaching the nosepiece
Positioning pin
Attachment tool
(for tightening the
nosepiece fixing
screw)
Positioning
groove