4
Attach a tube.
Note on device combinations
A P-TERG100 or P-TERG50 trinocular tube cannot be used in the microscope system that includes any of the following
stands:
 C−PSN/C-PSCN Plain Stand
 C-DS Dia Illumination Stand
 C-LEDS Hybrid LED Stand
Attach a tube to the zooming body. The procedure for
attaching a tube is the same for all types.
(1)
Completely loosen the M4 set screw for the tube at
the upper front of the zooming body using a hex
wrench (nominal designation: 2).
(2)
Align the positioning groove on the rear of the round
dovetail at the bottom of the tube with the positioning
pin on the rear of the round dovetail groove at the
zooming body upper plane so that they fit, and then
tighten the fixing screw.
5
Attach eyepieces.
Attach eyepieces to the tube.
Use two eyepieces of the same magnification. Completely
insert them into the tube until they touch the edge plane of
the tube sleeve.
Caution when attaching the eyepieces
Pay attention when inserting a 10x eyepiece because a rubber cover is attached to the 10x eyepiece so the eyepiece
end is not visible. To prevent malfunction, hold the rubber cap when inserting the 10x eyepiece. Do not hold the diopter
adjustment ring when inserting it.
Chapter 5
Assembly
69
Securely attaching the tube
Attaching the eyepieces
Positioning
pin