Nikon SMZ800N Instruction Manual page 80

Stereo microscope
Table of Contents

Advertisement

Beware of overhang of cables
When these cables are connected, the connectors (cables) overhang to the left of the base. Be careful so that they do
not interfere with other devices such as a control box or they are not accidentally touched and disconnected while
attaching devices or performing observation.
Use the power cord specified in Chapter 8, "2 Performance Properties."
WARNING
Before connecting the power cord
To prevent electric shock, be sure to turn off the power switch (press it to the "O" position) at the front of the base before
connecting the power cord to the adapter.
 Fiber attachment to the P-DSF32 Fiber Diascopic
Illumination Stand
Insert a hex driver (nominal designation: 2) in the fiber
fixing screw hole at the right rear side of the base and
loosen the internal fixing screw (M4 set screw). Insert the
fiber to the rear side of the base until it reaches the limit
and tighten the fixing screw.
Firmly insert the other end of the fiber into the fiber
attachment hole of the C-FLED2 LED Light Source for
Fiber Illuminator. See the instruction manual for the light
source for details.
 C-PSN/C-PSCN Plain Stand's arm position change
The arm of the C-PSN/C-PSCN Plain Stand can be
attached to a lower position.
Loosen the arm attachment screw and attach the arm to
the lower screw hole position on the vertical slider using
the hex wrench (nominal designation: 4) supplied with the
stand. The arm can be attached to the position 55 mm
lower than the standard. Check that two pins of the arm
are fitted in the groove of the vertical slider before
tightening the fixing screw.
Chapter 5
Assembly
66
Securely attaching the fiber
Groove
Pins
Arm position change