Perforated Plate; Fig. 11, Replacement Of The Perforated Plate - Heraeus HS 9 Service Manual

Microbiology safety cabinets
Table of Contents

Advertisement

Service Instructions HERAsafe HS/12/18
4.
Replacement of defined spare parts
4.6

Perforated plate

Replacement of the perforated plate:
NOTE - Safety precautions:
Observe the safety instructions!
Check decontamination of the unit!
Removal:
1. Open the front window [1] as wide as possible.
2. Shut the unit down.
3. Unscrew the two screws [2] .
4. Remove the perforated plate [3] from the work
chamber.
Installation:
5. Place the perforated plate [2] on the rear seating
with the beads facing upward.
6. Push the perforated plate backward, then place it
on the front seating.
NOTE - Unit commissioning:
Prior to the commissioning of the unit:
Perform test run!
Check safety devices for proper opera-
tion!
Valid: 05.2001 / 50060091 / A
Figure 11

Fig. 11, Replacement of the perforated plate

• 23 •