Test Report Clean Air Cabinets
Customer
Model
Order No.
Operation hours
Test according to VDI 2083 part 3 and FED 209 E
Instruction Manual ?
Log book ?
Type plate ?
Clean Room Class
Modulwidth
Remark:
3. Electrical safety in accordance to IEC 1010
Test Equipment
Bonding impedance
Isolation impedance
Leakage curretn
Test Equipment
No optical leakages
Filter sealing
Filter class (efficiency 0.3 µm)
No. of particles before filter 0.3µm
Max. allowed mean
penetration
Max. allowed local
penetration
Meet all requirements
Remark:
Test equipment
Anemometer
Measure plane
300 mm in front of filter surface
Test location
Serial - No.
Date
1. Visual Inspection
Operating instruction ?
Latest test report ?
Safety markings ?
2. Generall Inspection
Moduldepth
[m]
Safety test Equipment
< 0,2
> 2 M /500 V DC
< 7mA
4. Filtertest in accordance to VDI 2083
Laser Particle counter 0.3µm, aerosol dilution device
1:1000, aerosole generator, test aerosol DEHS
Measured mean
penetration
Measured local
penetration
Yes
5. Air velocity
Page 1
measured
No
VDI 2083_FED 209_engl.xls