Heraeus HS 9 Service Manual page 174

Microbiology safety cabinets
Table of Contents

Advertisement

Manufacturing site./Year
Billing Address Customer no.:
Manufacturing year:
Customer center:
Order date:
Ordered by:
Equipment type:
Equipment no.:
Serial no.:
st
1
page
Testing Safety Cabinet Class II according to EN 12 469
Installation test
Laboratory Class
Safety Declaration filed
Visual check
Type label
GS – label
Brief operating instructions
Warnings
Attach test label
Check joints of sealing agents
Check surface for ruptures
Safety check
Ground conductor
Insulating resistance
Monitoring equipment, acoustic and visual
Contactor installed in the instrument
Contactor for upper limit responds
Contactor for lower limit responds
Front panel position - Alarm triggers if front panel outside working position
Remarks:
Order no.:
Technician no:
Technician's Name:
Order no.:
Service reference no.:
Tested
Remarks
Values
OK
VDE 0701-1 :2000-09
< 0,3
> 2 M
Sheet:
Installation address
Part no.:
Delivery date:
Installation date:
Routine maintenance test
Complies with installation rules
Yes :
No:
Visual check
Direct with exhaust
DD
air trap
TD
Draught interrupt
KÜHLER Exhaust cooler
nd
2
exhaust air filter
AEF
Check suction protection,
exchange if necessary
Height of inflow opening
Function monitor contact
(Potential-free contact)
Safety check
Conductance
Test Record
Date:
Hours of use:
Customer inventory no.:
Not required:
Tested
Remarks
Values
OK
VDE 0701-1 :2000-09
< 3.5 mA
Yes
No
Yes
No
Yes
No
Yes
No

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Hs 12Hs 15Hs 18Hsp 9Hsp 12Hsp 15 ... Show all