9.4. Printing rhythm events

The device is able to continuously print the rhythm events listed in the previous section in 4-
second intervals.
In order to print the rhythm events continuously, event print has to be enabled
(Setup/Printing/Event Print). This initialises the printing of arrhythmia events. A yellow
"E" character appearing to the left of the battery icon on the monitoring screen indicates that
event printing is active.
While the program is running, in case an event listed in the section above occurs, a 4 second
long curve is printed, with the event approximately at its centre. After starting the program,
the length of the first event is approx. 6 seconds, during which the patient data is printed the
same way as in the ECO printing mode. The following printouts are 4 seconds long, and
indicate the time of printing below the curves.
Caution!
The results of the measurements are not stored except on the printout!
When printing rhythm events is initialised, the device enables the pre-start option, ECO
printing, and turns the baseline filter off. Disabling any of these functions, turns ECO printing
off as well.

9.4.1. Aborting the program

Printing rhythm events can be stopped by pressing the
or
button, or by turning it
off in the menu.
R-6730-EN02-E
39/73