Innomed Medical HS112 C-1 User Manual page 73

12-channel ecg device with 112 mm wide built-in printer, colour touch screen lcd display
Table of Contents

Advertisement

Error /
Error message
The device is printing,
but there is nothing on
the paper.
Printouts are faded.
After some time (a few
years)
the
printout
fades.
The operation time of
the device running on
battery is limited.
The
device
connected to the mains
power supply, but the
charge indicator LED
does not light up.
R-6730-EN02-E
HS112 C-1 User Manual
Possible cause
The upper side of the
paper is not
thermosensitive.
Paper sensitivity is
too low.
Paper-related issue.
The battery is too
old.
is
The battery is fully
charged.
Solution
The paper is either unsuitable, or it has
been
loaded
upside
thermosensitive side should face upwards.
Use the paper recommended by the
manufacturer.
Use the paper recommended by the
manufacturer. With proper storage, the
content is preserved for up to 8 years.
Make a photocopy or scan of the printout.
Replace the battery.
If the battery is fully charged, the LED
should not light up. However, if the battery
is not charged, check the fuse, or contact
the service.
down
––
the
73/73