3.3. Battery related tasks

Before first use, the battery has to be charged. In order to do this, connect the device to a
mains power outlet. Make sure to take into consideration the warnings listed in section 1.
The LED light labelled "Charge" located on the front panel of the device indicates that the
battery is charging.
The device battery can be charged continuously without damage; however, it is advised not to
leave the device charging overnight or unattended. In these cases, the charging process can be
interrupted by disconnecting the device from the mains power outlet.
In case there is an interruption in the power supply during charge, and there is a charged and
working battery in the device, the charging process will stop, but the device will continue to
work uninterrupted until the battery is discharged. Stored data will not be lost.
In case of a defective or insufficiently charged battery, the device will shut down if there is an
interruption of power supply. When turning the device back on, the screen will be black.

3.4. Environmental protection

3.5. Classification

The HS112 C-1 electrocardiograph is classified as a Class IIa device according to Rule 10 of
point 3.2 of Part III of Annex IX to Council Directive 93/42/EEC as modified by Directive
2007/47/EC.
With respect to electromagnetic compatibility, the HS112 C-1 electrocardiograph is classified
as Group 1, Class B according to section 4 of the MSZ EN 55011:2000 standard.
R-6730-EN02-E
HS112 C-1 User Manual
Disposal of old electrical and electronic
equipment (applicable in the European Union
and other European countries with separate
collection systems)
This symbol on the product or on its packaging
indicates that this product shall not be treated as
household waste. Please hand it over to the
applicable collection point for the recycling of
electrical and electronic equipment.