Innomed Medical HS112 C-1 User Manual page 35

12-channel ecg device with 112 mm wide built-in printer, colour touch screen lcd display
Table of Contents

Advertisement

 The program stores the time and type of the R-waves, and calculates a range of data and
statistics based on these when the recording is finished. The measured time period
corresponds to the pre-set rhythm period, but measuring can be stopped at any time by
pressing the STORE button. If the program has detected at least 40 beats before it was
aborted, the analysis is performed; however, if there were less than 40 beats, no analysis
is performed.
The arrhythmia program examines the heart rhythm and QRS width, and classifies the beats
based on them. For rhythm analysis, the program utilizes a specialized algorithm to calculate
the average R–R distance, and compare the current R–R value to it. Rhythm-based categories
are:
Normal
Post-extra
Extra
Long
QRS width is measured on 3 channels by determining the beginning and the end of the QRS
on each of them, and calculating the QRS width based on these measures. Width-based
categories are:
Normal
Wide
Extra wide
In the case of a long beat, the program does not differentiate between beats based on their
width. It will indicate that there has been a change at the stimulus source (e.g. sinus
arrhythmia or escape beat) on all 3 channels.
Possible events by categories:
 Wide, extra wide rhythmic beats:
Left bundle branch block beat
Right bundle branch block beat
Left or right bundle branch block beat
Fusion of ventricular and normal beat
Ventricular escape beat
 Wide, extra wide premature beats:
Aberrant premature atrial beat
Premature ventricular contraction
R on T
 Wide, extra wide long beats:
Left or right bundle branch block beat
Ventricular escape beat
 Premature narrow beat:
Nodal/junctional/premature beat
Atrial premature contraction
Premature or ectopic supraventricular beat
R-6730-EN02-E
HS112 C-1 User Manual
QRS_length < 120 ms
120 ms < QRS_length < 160 ms
160 ms < QRS_length
LBBB
RBBB
BBB (wide)
FUSION
VESC (extra wide)
ABERR
PVC (wide, extra wide)
RONT
BBB
VESC
NPC
APC
SVPB *
35/73

Advertisement

Table of Contents
loading

This manual is also suitable for:

Heartscreen112 clinic

Table of Contents