11.2. Deleting a record

The selected record can be deleted by pressing the DEL (AUTO) button. The device will
request a confirmation by asking: "Are you sure you want to delete?"
"No: AUTO Yes: LEAD" will appear at the bottom of the display. If the LEAD button is
pressed, the record is deleted, and the catalogue without the deleted record is shown on the
display. If the AUTO button is pressed, the record remains in the catalogue.
It is not necessary to delete records manually as the device deletes the oldest record
automatically in case the memory is full.

11.3. Printing all record data

Pressing the
displayed prints out all stored elements of the selected record (indicated by the red record
number). The elements are printed in the following order — if available:
Diagnosis program
 Header
 Stored ECG record of the 12 channels
 Rhythm curve (with a 5 mm/s sweep speed)
 Average curves (with a fixed 25 mm/s sweep speed)
 Analysis data table
 Diagnosis (Stored only if a diagnosis was successfully created.)
Store program
 Header
 Stored ECG record of the 12 channels
 Rhythm curve (with a 5 mm/s sweep speed)
Store program with arrhythmia analysis
 Header
 Stored ECG record of the 12 channels
 Rhythm curve
 Arrhythmia summary data
 Arrhythmia diagrams
Printing can be stopped by pressing the
R-6730-EN02-E
HS112 C-1 User Manual
button in the main menu of the catalogue when the list of records is
button one more time.