5. Installing the HS112 C-1 device

The device is shipped in a carrying case which contains the device, accessories, and the user
manual. Perform the following steps after unpacking the device.
Before turning on the device:
 Read the electrocardiograph's manual and familiarize yourself with the layout and
functionality of the controls.
 Make sure that the device and the cables are not damaged.
 Connect the patient cable to the HS112 C-1 device.
 Load the paper into the ECG device as specified in section 5.2.
 Connect the device to the mains power supply to charge the battery.
Caution!
The device is only allowed to be connected to a power supply in accordance with the
specifications. The only way to completely disconnect the device from the mains power source
is to pull the power plug out of the mains outlet. Install and position the device in a way that
makes this possible.

5.1. Battery-related tasks

 The device continuously keeps track of the charge status of the battery, and in case the
device is operating on battery, a battery indicator symbol appears in the top right
corner of the display under the HR symbols. When the device operates on the mains
power supply, there is no battery indicator. If the charge level of the battery is very
low, the indicator turns red and starts to blink (in the rhythm of the QRS). The colour
of the battery indicator reflects the charge level; however, it does not directly
correspond to the remaining operational time.
 If the device does not turn on, it has to be connected to the mains power supply in
order to charge.
 The device battery can be charged continuously without damage; however, it is
advised not to leave the device charging overnight or unattended. In these cases, the
charging process can be interrupted by disconnecting the device from the mains power
outlet.
 If the device is not being used (it is not connected to a patient and/or no button is
pressed for 5 minutes), and if "Turn off" is set to 5 minutes in the Setup/Others
menu, the device will shut down automatically after approximately 5 minutes.
However, the battery will continue to charge.
 The LED light labelled "Charge", which indicates that the battery is charging, is
located at the bottom of the front panel, next to the power button.
 To preserve battery capacity, it is recommended to format the battery at least every 6
months.
R-6730-EN02-E
HS112 C-1 User Manual
13/73