Digitalizing Features; Printing Features; Display - Innomed Medical HS112 C-1 User Manual

12-channel ecg device with 112 mm wide built-in printer, colour touch screen lcd display
Table of Contents

Advertisement

Time constant
Patient leakage current
Electrode downfall detection
Calibrating signal (software generated)
Muscle filter (digital)
Line filter (digital)
Line filter - 20 dB point:
Baseline filter (digital)
Amplifier error messages
Patient cable type
Innomed approved patient cable
Defibrillator tolerance
Only with defibrillator protected cable!

18.10. Digitalizing features

Data acquisition
Sampling frequency
A/D resolution
Nominal processed signal range
Stored signal resolution
Resolution in case of PC connection during online ECG
transfer
Recording Format

18.11. Printing features

Effective recording width per channel
Printer head resolution
Printing speed
Paper width
Paper block (folding)
Paper replacement
Warnings
Printed data
Printed analysis data

18.12. Display

Display type
Resolution
Number of displayed curves
R-6730-EN02-E
HS112 C-1 User Manual
Date, time, patient identifier code, age, speed,
sensitivity, lead marks, filter status
Amplitude and time parameters measured on averaged
QRS cycle, heart frequency, axis positions and Q-T,
P-Q, QRS width and diagnostic categories
The calculated time parameters are signed by markers
on the average curves
50 Hz filter:
48,5 ... 51,5 Hz
60 Hz filter:
59 ... 61 Hz
complete suppression
under 0,3 Hz:
complete transfer
over 0,9 Hz:
Audio and visual signal in case of
input level error and noisy signal
Defibrillator protected 10-lead cable
max. 360 J, 5000 V
12-lead, simultaneous (within 100us)
8000 Hz / channel
24 bit, 0.076uV/bit
500 Hz 10 uV/bit
500 Hz 5 V/bit
40 mm, overlap allowed
In amplitude direction: 8 points / mm
On the time axis: 8 or 16 points / mm
112 mm +0 -1 mm
300 pages, 150 mm long pages
Sound signal when out of paper
TFT colour, LCD touch screen, size: 5,7"
3 channels + with rhythm lead group change
6 channels + with rhythm lead group change
12 (2 x 6) channels + rhythm
min. 3,2 sec
max. 10 A
by channels
1 mV / 300 ms
35 Hz +0/-3 dB
50 / 60 Hz
4-lead cable
700 mV
PDF
5, 25, 50 mm/s
automatic feed
640 x 480 points
61/73

Advertisement

Table of Contents
loading

This manual is also suitable for:

Heartscreen112 clinic

Table of Contents