1. Introduction, warnings

For safe operation, please comply with the following warnings:
HS 112 C-1 is a safety class I ECG device with type CF patient isolation and defibrillator protection.
!
The mains cable of the device can be connected to a grounded power outlet only.
If the mains power outlet is damaged or not working, or the grounding is inadequate, the device may only
!
be operated using the built-in battery!
!
Only the type of fuse specified in the documentation is allowed to be used in the device.
The use of paper recommended by the manufacturer is advised, as only this ensures the printer head
!
lifetime indicated in the technical specifications.
At most 10 electrodes may be connected to the patient. The use of separate grounding is FORBIDDEN!
!
Make sure that the unused electrodes (including the N electrode) do not come in contact with any current-
carrying objects.
When connecting the device to a measuring system, make sure that only appropriately isolated and IEC
!
60601 compliant devices are connected, and that the patient does not come into contact with any devices
connected to the mains power supply.
If the device is used connected to a computer, make sure that the patient is kept at an at least 1.5 m
!
distance from the computer, or that the computer is operated from an isolation transformer.
The defibrillator protection of the device is guaranteed only with one of the following defibrillator
!
protected cables approved by the manufacturer.
Cable type:
The ECG evaluation program suggests a suitable diagnosis. The final decision is the physician's
!
responsibility!
Do not use the device in the presence of combustible, potentially explosive gases.
!
Disassembly or repair of the device is allowed only to experts trained by the manufacturer.
!
In case of hot environmental temperatures, the electrodes should be 41°C max when they are placed on the
!
patient.
The device cannot be used for direct, invasive cardiological measurement!
!
In the interest of protecting the environment, please adhere to the following:
When replacing the battery, the used or damaged battery should be disposed of only at a designated place for the collection of
hazardous waste (NiMH batteries).